Overview
MicroECG is an ECG monitoring module based on nRF51822 BLE chip and BL1860 ECG chip, which can transport heart information through either UART or BLE. Users can control nRF51822 via AT Commands. Also, they can use the open-source software to develop their own firmware, making it to support more functions.
The heart information includes row data and the heart rate. The row data is the result of ADC detection, representing the strength of heart beat. The row data can be used to draw ECG directly. And the heart rate is the result of calculation with the row data.
This ECG module is easy to use. Data can be transported through either UART or BLE. When the switch on the board is set to the BLE side, data can be transported through BLE. In this case, one can use app on one’s phone to receive the heart rate from this module. When the switch on the board is set to the UART side, data can be transported through UART. In this case, one can control the board with some AT commands.
Go shopping MicroECG (SKU:IM161018001) (Coming Soon!)
Specification
PCB Size | 22mm X 26mm X 1.6mm |
---|---|
Input Voltage (vcc) | 3.3V |
Microprocessor | nRF51822 |
Interface | UART, BLE |
Electrical Characteristics
Parameter | Min | Typical | Max | Unit |
---|---|---|---|---|
Input Volt | 3 | 3.3 | 3.6 | VDC |
Input voltage VinH | 3 | 3.3 | 3.6 | V |
Input voltage VinL | -0.3 | 0 | 0.5 | V |
Baud rate | 9600 | bps |
Hardware
J7 | |
---|---|
No. | Pin Name |
1 | 3V3 |
2 | SWDCLK |
3 | SWDIO/RST |
4 | TXD(P0.12) |
5 | RXD(P0.11) |
6 | GND |
7 | GND |
8 | LA |
9 | RA |
10 | RLD |
11 | NC |
12 | NC |
Interface Functions
- SWDCLK: SWD interface clock line
- SWDIO/RST: SWD interface data line
- RXD/TXD: serial, support AT command
- LA: left arm signal input pin
- RA: right arm signal input pin
- RLD: right arm drive feedback output pin
1.Power up this module with 3.3V. And use ECG Limb Clamp (IM161025001) and ECG Lead Wire (IM161025002) to clamp your arm.
2.Run the BLE app on the phone. Click and turn into the “HRM” mode and then connect to the “ITEAD_BL18_ECG” device.
3.Just wait a while and the heart rate will be displayed.
AT commands:
1.“AT\r”
This command is a string and used to test the UART connection.
When the UART is working , “OK\r\n” will be returned.
2.“ATdataTypeR\r”
This command is a string and used to set the data type to row data.
When the command runs successfully , "Type:R\nOK\r\n" will be returned.
This command can not run successfully when the “ATstartRead\r” is running.
3.“ATdataTypeH\r”
This command is a string and used to set the data type to heart rate.
When this command runs successfully, "Type:H\nOK\r\n" will be returned.
This command can not run successfully when the “ATstartRead\r” is running.
4.“ATdataType?\r” This command is a string and used to get the current data type.
When this command runs successfully, and the current data type is row data, then "Type:R\nOK\r\n" will be returned.
When this command runs successfully, and the current data type is heart rate, then "Type:H\nOK\r\n" will be returned.
This command can not run successfully when the “ATstartRead\r” is running.
5.“ATtestCon\r”
This command is a string and used to test the connection of the ECG cable.
When the hardware connection of ECG cable is ok, “OK\r\n” will be returned.
When the hardware connection of ECG cable is not ok, “ERROR\r\n” wii be returned.
6.“ATstartRead\r”
This command is a string and used to get the data from UART.
When the hardware connection of ECG cable is not ok, “ERROR\r\n” wiil be returned.
When the hardware is ok, “OK\r\n” will be returned firstly, and then the data will be transported continuously as soon as the data itself is updated.
7.“ATstopRead\r”
This command is a string and used to stop the reading action through UART.
When this command runs successfully, “OK\n\r” will be returned and the reading action will stop.